    <u id="gnb5o"></u>
  • Hemosiderin Stain

    Release time:2012-5-14      Reads:2049

    Hemosiderin Stain (Potassium Ferrocyanide)

     

     

    Intended use:

    Potassium ferrocyanide is intended for staining hemosiderin and differentiating hemosiderin from other pigments.

     

    Principle:

    The iron from tissue (in the ferric state) can be released from protein with dilute hydrochloric acid. It then reacts with potassium ferrocyanide to form blue ferric ferrocyanide, which is known as the Prussian blue reaction.

     

    Methods:

    1. Fix tissues in 10% neutral buffer formaldehyde. Dehydrate in typical manner and embed.

    2. Deparaffinize tissue in distilled water. Rinse with distilled water for another time.

    3. Mix an equal volume of 2% Potassium ferrocyanide and 2% hydrochloric acid. Drop the mixture onto slides for 20~30 minutes. Rinse with distilled water.

    4. Counterstain in Nuclear Fast Red solution for 5~10 minutes and rinse briefly with distilled water.

    5. Dehydrate in 95% ethanol and absolute ethanol, clear in xylene, and mount with mounting media.

     

    Specifications:

     

    Contents

    3vialsx20ml/kit

    3Btlsx100ml/kit

    3Btlsx250ml/kit

    Components

    Potassium ferrocyanide solution

    20ml

    100ml

    250ml

    Potassium ferrocyanide

    hydrochloric acid aqueous solution

    20ml

    100ml

    250ml

    Hydrochloric acid

    Nuclear Fast Red solution

    20ml

    100ml

    250ml

    Nuclear Fast Red

     

     

    Precaution:

    1. Tissues should be fixed with neutral buffered formaldehyde. If ordinary formaldehyde is used, the tissue should be fixed for a shorter time before embedding. Tissues that are fixed in ordinary formaldehyde for a long time react poorly. Avoid using fixatives that contains chromate.

    2. Hydrochloric acid of analytical grade or better is preferred as lower grade of hydrochloric acid typically contains iron impurities that can lead to false positive.

    3. All containers/utensils used throughout the entire procedure should be clean. Avoid using any iron tools.

    4. For steps prior to iron reaction, rinse only with distilled water to prevent iron in tap water from compounding with calcium salt in tissues that leads to false positive.

     

    Expected Results:

     

    Hemosiderin

    ——

    Blue

    Nuclei

    ——

    Red

     

     

    Back>>Top